16827 Mercury SCREW, VERTICAL STARTING DEVICE


16827 SCREW, VERTICAL STARTING DEVICE Mercury 1002201FM, 1002201JK, 1002201VB, 1003201NK, 1004137, 1004201FB, 1004201JK, 1004201VB, 1006201AK, 1006201FK, 1008201YM, 1025207FL, 1025207FM, 1025217FL, 1025217FM, 1030201FL, 1030201YL, 1030211FL, 1030271HL, 1030302FL, 1030312FL, 1030A01FL, 1030A11FL, SCREW
16827 SCREW, VERTICAL STARTING DEVICE Mercury
Rating:
48

Buy SCREW, VERTICAL STARTING DEVICE 16827 Mercury genuine, new aftermarket parts with delivery
Number on catalog scheme: 6
 

Mercury entire parts catalog list:

1002201FM 2006
1002201JK 1990,1991,1992,1993,1994,1995,1996,1997,1998
1002201VB 1999,2000,2001,2002,2003,2004,2005
1003201NK 1993,1994,1996,1997,1998
1004137 1987,1988,1989
1004201FB 2006
1004201JK 1990,1991,1992,1993,1994,1995,1996,1997,1998
1004201VB 1999,2000,2001,2002,2003,2004,2005
1006201AK 2003,2004
1006201FK 2006
1008201YM 2001,2002,2003,2004
1025207FL 2006
1025207FM 2006
1025217FL 2006
1025217FM 2006,2010
1030201FL 2006
1030201YL 2001,2002,2003,2004
1030211FL 2006
1030271HL 2006
1030302FL 2006
1030312FL 2006
1030A01FL 2006
1030A11FL 2006
1031207ZF 2002
1B04201FB 2006
1F04201WB 2000,2001,2002,2003,2004,2005
1F25A17DM 2005

Information:

Required Equipment
The following is a list of equipment required to test and calibrate temperature contactors.* 6V-7070 Multimeter or Equivalent A multimeter is used to verify the actuation and deactuation settings. This is done by checking for continuity between two terminals inside the contactor.* Heat Bath A heat bath is used to maintain a constant elevated temperature with the temperature probe submersed into a silicone heating fluid. This allows for the calibration of the actuation and deactuation settings. It must be temperature controlled and meet the following specifications:a. Capable of obtaining a temperature of 150° C (302° F) or have at least a 20° C (40° F) higher temperature capacity than the specified set point.b. Maintain the temperature within 1° C ( 2° F) maximum temperature difference.c. Must be equipped with a thermometer or digital display temperature read-out accurate to within 0.5° C ( 1° F).d. Circulate the heating fluid to maintain a uniform temperature.e. Must have a minimum bath capacity of 0.5 L (31.25 cu in). Also requires a working depth large enough to fully immerse the temperature probe.f. Must be compatible with silicone fluids.g. Can be a portable self-contained unit. The heat baths listed in Chart B meet the specified requirements and can be purchased from:Cole-Parmer Instruments Co.
7425 North Oakpark Ave.
Niles, IL, 60714 U.S.A.
Phone (800) 323-4340
Fax: (708) 647-9660
* Heat Bath Fluid The heat bath fluid must be able to obtain and maintain the required contactor actuation and deactuation temperatures. For the temperatures required for the calibration procedure, silicone fluid must be used. Silicone fluid has a temperature range between 10° C (50° F) and 230° C (446° F). Silicone fluid can be purchased from Cole-Parmer by ordering G-01294-40. The fluid comes in a 3.8 Liter (1 gal) container.
Use extreme caution around heat baths with high temperature fluids. High temperature fluids can cause severe burns.
Calibration Procedure
Use the following procedure to test and calibrate the temperature contactors.1. If connected, disconnect the contactor from the power supply. Mark the terminal location and disconnect the three wires from terminals 1, 2, and 4.
High voltage electrical shock is possible. The contactor must be disconnected from the power supply and the wires disconnected from terminals 1, 2, and 4. Failure to follow this recommendation can result in serious bodily injury.
2. Locate and record the part number of the temperature contactor in Chart A. This will determine the actuation and deactuation temperatures. If the temperature contactor is used in extreme ambient temperatures there will be a scale deviation from the required actuation and deactuation temperatures. Refer to the "Calculating Scale Correction for Severe Temperature Applications" section in this Special Instruction.
Temperature Contactor
(1) Housing cover. (2) Housing cover screws. (3) Temperature probe. (4) Capillary tube.3. Loosen housing cover screws (2) and remove housing cover (1).4. Fill the heat bath unit with silicone fluid and heat it to approximately 5° C (10° F) below the actuation temperature setting. Do not immerse the temperature probe at this time.
Nomenclature Of Internal Parts
(5). Lock screw. (6) Range spindle. (7) Range scale. (8) Differential spindle. (9) Differential scale.5. Loosen lock screw (5).6. Adjust range spindle (6) to a setting above the desired actuation setting determined in Step 2. Determine the setting by using range scale (7).7. Adjust the differential spindle (8) to the approximate mechanical differential temperature using differential scale (9). Mechanical differential temperature is the difference between the actuation and deactuation temperature settings.
Temperature Control Function For Actuation And Deactuation (specifically shown for 7C-3888 Contactor)
Adjusting Temperature and Differential Settings On The Temperature Contactor
This example shows the temperature set at 66° C with an 8° C differential setting. This results in a deactuation temperature of 58° C and actuation temperature of 66° C. (Refer to "Temperature Control Function For Actuation And Deactuation" illustration.)8. Fully immerse temperature probe (3) in the silicone fluid. Increase the heat bath temperature to the specified actuation temperature.9. When the actuation temperature has stabilized, very slowly turn range spindle (6) counterclockwise until the contact actuates. This will be accompanied by an audible "click".
When Contact Actuates, Terminals 1 And 4 Will Have Continuity10. Use a 6V-7070 Multimeter, or equivalent, to verify the actuation of the contactor. Check for continuity between terminals 1 and 4. If there is no continuity, the contactor has not actuated; continue to adjust the range spindle until there is continuity between terminals 1 and 4.11. Slowly decrease the heat bath temperature until the contact deactuates. Again, this will be accompanied by an audible "click".
When Contact Deactuates, Terminals 1 And 2 Will Have Continuity12. Verify the deactuation of the contactor by checking for continuity between terminals 1 and 2. If there is no continuity, the contactor has not deactuated; continue to decrease the temperature until it does. Record this deactuation temperature.13. Again, slowly increase the heat bath temperature until the contact actuates. This gives an accurate indication of the actual setting. If the actuation setting is not correct (too high or too low), adjust range spindle (6) until the required actuation setting is reached.14. Once the actuation setting is established, the deactuation setting can be slightly raised or lowered by adjusting differential spindle (8). Adjusting differential spindle (8) will generally not affect the actuation setting, but it is advisable to recheck it after any adjustments.15. Continue to adjust differential spindle (8) until the deactuation setting is established.16. Tighten lock screw (5) when both


Parts screw Mercury:

95386
 
95386 SCREW, PLATE TO TOP COWL
1002201FM, 1002201JK, 1002201VB, 1002204, 1003201NK, 1030201YL, 1031207ZF, 1040203FL, 1040203YL, 1040213YL
40011107
SCREW, HOUSING TO ADAPTOR PLATE (16MM)
40011107 SCREW, HOUSING TO ADAPTOR PLATE (16MM)
1006201DB, 1006201PD, 1006201RB, 1006201VB, 1006206, 1006211RD, 1008211RD, 1010207VB, 1011201DB, 1016207PD, 1016207RB, 1016207SB, 10202014D, 1020201BC, 1020201DB, 1020201VB, 1031203PD, 1031203UB, 1031203UD, 1031207ZF, 1031312DB, 1040213YL, 1041312UB,
16059
 
16059 SCREW, ALTERNATOR
1002201FM, 1002201JK, 1002201VB, 1003201NK, 1004137, 1004201FB, 1004201JK, 1004201VB, 1006201AK, 1006201FK, 1008201YM, 1F04201WB
16122
 
16122 SCREW (SHORT)
1004137, 1004201FB, 1004201JK, 1004201VB, 1030201YL, 1031203UB, 1031203UD, 1031207ZF, 1031312DB, 1040203FL, 1040203YL, 1040213YL, 1B04201FB, 1B08201DB, 1F04201WB
17173
 
17173 SCREW
1004201FB, 1004201JK, 1004201VB, 1B04201FB, 1F04201WB
803509 3
 
803509 3 SCREW, Drain (M5 x 6)
1B04201FB, 1F02201HM, 1F04201WB
878388
SCREW, Stainless Steel
878388 SCREW, Stainless Steel
1A25203FK, 1A25203FL, 1A25311FK, 1A25312FK, 1A30413KZ, 1B04201FB, 1B08201DB, 1F04201WB, 1F15211HM
40116110
SCREW, (M6 x 110)
40116110 SCREW, (M6 x 110)
1015203FL, 1015203YL, 1025207FL, 1025207FM, 1025217FL, 1025217FM, 1030201FL, 1030201YL, 1030211FL, 1030271HL, 1030302FL, 1030312FL, 1030A01FL, 1030A11FL, 1031207ZF, 1B04201FB, 1F04201WB
Back to top